Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide 345
Introduction to
:MEASure
Commands
The commands in the MEASure subsystem are used to make parametric
measurements on displayed waveforms.
Measurement Setup
To make a measurement, the portion of the waveform required for that
measurement must be displayed on the oscilloscope screen.
Measurement Error
If a measurement cannot be made (typically because the proper portion of the
waveform is not displayed), the value +9.9E+37 is returned for that measurement.
:MEASure:XMAX
[<source>] (see
page 387)
:MEASure:XMAX?
[<source>] (see
page 387)
<source> ::= {CHANnel<n> |
FUNCtion | FFT | MATH |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= horizontal
value of the maximum in NR3
format
:MEASure:XMIN
[<source>] (see
page 388)
:MEASure:XMIN?
[<source>] (see
page 388)
<source> ::= {CHANnel<n> |
FUNCtion | FFT | MATH |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= horizontal
value of the maximum in NR3
format
Table 63 :MEASure Commands Summary (continued)
Command Query Options and Query Returns
Measurement Type Portion of waveform that must be displayed
period, duty cycle, or frequency at least one complete cycle
pulse width the entire pulse
rise time rising edge, top and bottom of pulse
fall time falling edge, top and bottom of pulse