User Manual

Table Of Contents
:MARKer Commands 19
Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide 333
:MARKer:Y2Position
(see page 782)
Command Syntax
:MARKer:Y2Position <position> [suffix]
<position> ::= Y2 cursor position in NR3 format
<suffix> ::= {mV | V | dB}
If the :MARKer:MODE is not currently set to WAVeform (see ":MARKer:MODE" on
page 324), the :MARKer:Y1Position command:
•Sets :MARKer:MODE to MANual.
Sets the Y2 cursor position to the specified value.
Y cursor units are set by the :MARKer:YUNits command.
When the :MARKer:MODE is set to WAVeform, Y positions cannot be set.
Query Syntax
:MARKer:Y2Position?
The :MARKer:Y2Position? query returns current Y2 cursor position. This is
functionally equivalent to the obsolete :MEASure:VSTOp command/query.
Return Format
<position><NL>
<position> ::= Y2 cursor position in NR3 format
See Also "Introduction to :MARKer Commands" on page 322
":MARKer:MODE" on page 324
":MARKer:X1Y1source" on page 326
":MARKer:X2Y2source" on page 328
":MARKer:Y1Position" on page 332
":MARKer:YUNits" on page 335
":MEASure:VSTOp" on page 728
NOTE
If the front-panel cursors are off or are set to Binary or Hex Mode, the marker position values
are not defined and an error is generated. Make sure to set :MARKer:MODE to MANual or
WAVeform to put the cursors in the front-panel Normal mode.