User Manual

Table Of Contents
324 Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide
19 :MARKer Commands
:MARKer:MODE
(see page 782)
Command Syntax
:MARKer:MODE <mode>
<mode> ::= {OFF | MEASurement | MANual | WAVeform}
The :MARKer:MODE command sets the cursors mode:
OFF — removes the cursor information from the display.
MANual — enables manual placement of the X and Y cursors.
If the front-panel cursors are off, or are set to the front-panel Hex or Binary
mode, setting :MARKer:MODE MANual will put the cursors in the front-panel
Normal mode.
MEASurement — cursors track the most recent measurement.
Setting the mode to MEASurement sets the marker sources
(:MARKer:X1Y1source and :MARKer:X2Y2source) to the measurement source
(:MEASure:SOURce). Setting the measurement source remotely always sets the
marker sources.
WAVeform — the Y1 cursor tracks the voltage value at the X1 cursor of the
waveform specified by the X1Y1source, and the Y2 cursor does the same for the
X2 cursor and its X2Y2source.
Query Syntax
:MARKer:MODE?
The :MARKer:MODE? query returns the current cursors mode.
Return Format
<mode><NL>
<mode> ::= {OFF | MEAS | MAN | WAV}
See Also "Introduction to :MARKer Commands" on page 322
":MARKer:X1Y1source" on page 326
":MARKer:X2Y2source" on page 328
":MEASure:SOURce" on page 371
":MARKer:X1Position" on page 325
":MARKer:X2Position" on page 327
":MARKer:Y1Position" on page 332