:FUNCtion[:FFT]:CENTer
(see page 782)
Command Syntax
:FUNCtion[:FFT]:CENTer <frequency>
<frequency> ::= the current center frequency in NR3 format. The range
of legal values is from 0 Hz to 25 GHz.
The :FUNCtion[:FFT]:CENTer command sets the center frequency when FFT (Fast
Fourier Transform) is selected.
Query Syntax
:FUNCtion[:FFT]:CENTer?
The :FUNCtion[:FFT]:CENTer? query returns the current center frequency in Hertz.
Return Format
<frequency><NL>
<frequency> ::= the current center frequency in NR3 format. The range
of legal values is from 0 Hz to 25 GHz.
See Also "Introduction to :FUNCtion Commands" on page 289
":FUNCtion[:FFT]:SPAN" on page 293
":TIMebase:RANGe" on page 563
":TIMebase:SCALe" on page 565
NOTE
After a *RST (Reset) or :AUToscale command, the values returned by the
:FUNCtion[:FFT]:CENTer? and :FUNCtion:SPAN? queries depend on the current
:TIMebase:RANGe value. Once you change either the :FUNCtion[:FFT]:CENTer or
:FUNCtion:SPAN value, they no longer track the :TIMebase:RANGe value.