284 Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide
16 :FRANalysis Commands
:FRANalysis:WGEN:LOAD
(see page 782)
Command Syntax
:FRANalysis:WGEN:LOAD <impedance>
<impedance> ::= {ONEMeg | FIFTy}
The :FRANalysis:WGEN:LOAD command selects the expected output load
impedance.
The output impedance of the Gen Out BNC is fixed at 50 ohms. However, the
output load selection lets the waveform generator display the correct amplitude
and offset levels for the expected output load.
If the actual load impedance is different than the selected value, the displayed
amplitude and offset levels will be incorrect.
Query Syntax
:FRANalysis:WGEN:LOAD?
The :FRANalysis:WGEN:LOAD? query returns the current expected output load
impedance.
Return Format
<impedance><NL>
<impedance> ::= {ONEM | FIFT}
See Also ":FRANalysis:DATA" on page 273
":FRANalysis:ENABle" on page 274
":FRANalysis:FREQuency:STARt" on page 277
":FRANalysis:FREQuency:STOP" on page 278
":FRANalysis:RUN" on page 280
":FRANalysis:SOURce:INPut" on page 281
":FRANalysis:SOURce:OUTPut" on page 282
":FRANalysis:WGEN:VOLTage" on page 285