280 Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide
16 :FRANalysis Commands
:FRANalysis:RUN
(see page 782)
Command Syntax
:FRANalysis:RUN
The :FRANalysis:RUN command performs the Frequency Response Analysis. This
analysis controls the built-in waveform generator to sweep a sine wave across a
range of frequencies while measuring the input to and output from a device under
test (DUT). At each frequency, gain (A) and phase are measured and plotted on a
Bode frequency response chart.
The :FRANalysis:APPLy command is a valid compatible alias for the
:FRANalysis:RUN command.
When the frequency response analysis completes, you can use the
:FRANalysis:DATA? query to get four comma-separated columns of data for each
step in the sweep: Frequency (Hz), Amplitude (Vpp), Gain (dB), and Phase (°).
It takes some time for the frequency sweep analysis to complete. You can query bit
0 of the Standard Event Status Register (*ESR?) to find out when the analysis is
complete.
See Also ":FRANalysis:DATA" on page 273
":FRANalysis:ENABle" on page 274
":FRANalysis:FREQuency:MODE" on page 275
":FRANalysis:FREQuency:SINGle" on page 276
":FRANalysis:FREQuency:STARt" on page 277
":FRANalysis:FREQuency:STOP" on page 278
":FRANalysis:SOURce:INPut" on page 281
":FRANalysis:SOURce:OUTPut" on page 282
":FRANalysis:TRACe" on page 283
":FRANalysis:WGEN:LOAD" on page 284
":FRANalysis:WGEN:VOLTage" on page 285
"*ESR (Standard Event Status Register)" on page 114