:FRANalysis Commands 16
Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide 277
:FRANalysis:FREQuency:STARt
(see page 782)
Command Syntax
:FRANalysis:FREQuency:STARt <value>[suffix]
<value> ::= {10 | 100 | 1000 | 10000 | 100000 | 1000000 | 10000000}
[suffix] ::= {Hz | kHz| MHz}
The :FRANalysis:FREQuency:STARt command command sets the frequency sweep
start value. The frequency response analysis is displayed on a log scale Bode plot,
so you can select from decade values.
Query Syntax
:FRANalysis:FREQuency:STARt?
The :FRANalysis:FREQuency:STARt? query returns the frequency sweep start
setting.
Return Format
<value><NL>
<value> ::= {10 | 100 | 1000 | 10000 | 100000 | 1000000 | 10000000}
See Also ":FRANalysis:DATA" on page 273
":FRANalysis:ENABle" on page 274
":FRANalysis:FREQuency:STOP" on page 278
":FRANalysis:PPDecade" on page 279
":FRANalysis:FREQuency:MODE" on page 275
":FRANalysis:RUN" on page 280
":FRANalysis:SOURce:INPut" on page 281
":FRANalysis:SOURce:OUTPut" on page 282
":FRANalysis:WGEN:LOAD" on page 284
":FRANalysis:WGEN:VOLTage" on page 285