User Manual

Table Of Contents
:FRANalysis Commands 16
Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide 275
:FRANalysis:FREQuency:MODE
(see page 782)
Command Syntax
:FRANalysis:FREQuency:MODE <setting>
<setting> ::= {SWEep | SINGle}
The :FRANalysis:FREQuency:MODE command lets you select between the normal
swept frequency response analysis or analysis at a single frequency, which can be
useful when debugging.
Query Syntax
:FRANalysis:FREQuency:MODE?
The :FRANalysis:FREQuency:MODE? query returns the frequency mode setting.
Return Format
<setting><NL>
<setting> ::= {SWEep | SINGle}
See Also ":FRANalysis:RUN" on page 280
":FRANalysis:FREQuency:SINGle" on page 276
":FRANalysis:FREQuency:STARt" on page 277
":FRANalysis:FREQuency:STOP" on page 278
":FRANalysis:PPDecade" on page 279
NOTE
The SINGle mode is available through the remote programming interface only and is not
available when using the front panel graphical user interface (GUI).