266 Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide
15 :FFT Commands
:FFT:SPAN
(see page 782)
Command Syntax
:FFT:SPAN <span>
<span> ::= the current frequency span in NR3 format. Legal values are
1 Hz to 100 GHz.
If you set the frequency span to a value outside of the legal range, the
step size is automatically set to the nearest legal value.
The :FFT:SPAN command sets the frequency span of the display (left graticule to
right graticule) when FFT (Fast Fourier Transform) is selected.
Query Syntax
:FFT:SPAN?
The :FFT:SPAN? query returns the current frequency span in Hertz.
Return Format
<span><NL>
<span> ::= the current frequency span in NR3 format. Legal values are 1
Hz to 100 GHz.
See Also ":FFT:CENTer" on page 259
":FFT:DISPlay" on page 260
":FFT:OFFSet" on page 261
":FFT:RANGe" on page 262
":FFT:REFerence" on page 263
":FFT:SCALe" on page 264
":FFT:SOURce1" on page 265
":FFT:VTYPe" on page 267
":FFT:WINDow" on page 268
NOTE
After a *RST (Reset) or :AUToscale command, the values returned by the :FFT:CENTer? and
:FFT:SPAN? queries depend on the current :TIMebase:RANGe value. Once you change either