User Manual

Table Of Contents
:CALibrate Commands 9
Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide 193
:CALibrate:OUTPut
(see page 782)
Command Syntax
:CALibrate:OUTPut <signal>
<signal> ::= {TRIGgers | MASK | OFF}
The CALibrate:OUTPut command sets the signal that is available on the Gen Out
BNC:
TRIGgers — pulse when a trigger event occurs.
MASK — signal from mask test indicating a failure.
OFF — either no signal or the waveform generator output signal when
:WGEN:OUTPut is ON.
Query Syntax
:CALibrate:OUTPut?
The :CALibrate:OUTPut query returns the Gen Out BNC signal selection.
Return Format
<signal><NL>
<signal> ::= {TRIG | MASK | OFF}
See Also "Introduction to :CALibrate Commands" on page 190
":WGEN:OUTPut" on page 677