182 Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide
8 :ACQuire Commands
:ACQuire:SEGMented:COUNt
(see page 782)
Command Syntax
:ACQuire:SEGMented:COUNt <count>
<count> ::= an integer from 2 to 50 (w/100K memory) in NR1 format
The :ACQuire:SEGMented:COUNt command sets the number of memory segments
to acquire.
The segmented memory acquisition mode is enabled with the :ACQuire:MODE
command, and data is acquired using the :DIGitize, :SINGle, or :RUN commands.
The number of memory segments in the current acquisition is returned by the
:WAVeform:SEGMented:COUNt? query.
The maximum number of segments may be limited by the memory depth of your
oscilloscope. For example, an oscilloscope with 100K memory allows a maximum
of 50 segments.
Query Syntax
:ACQuire:SEGMented:COUNt?
The :ACQuire:SEGMented:COUNt? query returns the current count setting.
Return Format
<count><NL>
<count> ::= an integer from 2 to 50 (w/100K memory) in NR1 format
See Also ":ACQuire:MODE" on page 179
":DIGitize" on page 143
":SINGle" on page 160
":RUN" on page 158
":WAVeform:SEGMented:COUNt" on page 638
":ACQuire:SEGMented:ANALyze" on page 181
"Introduction to :ACQuire Commands" on page 175
Example Code "Example Code" on page 183
NOTE
Segmented memory is available on the DSOX1000-Series oscilloscope models that have the
SGM license.