User Manual

Table Of Contents
:ACQuire Commands 8
Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide 181
:ACQuire:SEGMented:ANALyze
(see page 782)
Command Syntax
:ACQuire:SEGMented:ANALyze
This command calculates measurement statistics and/or infinite persistence over
all segments that have been acquired. It corresponds to the front panel Analyze
Segments softkey which appears in the Segmented Memory menu.
In order to use this command, the oscilloscope must be stopped and in segmented
acquisition mode, with either quick measurements or infinite persistence on.
See Also ":ACQuire:MODE" on page 179
":ACQuire:SEGMented:COUNt" on page 182
"Introduction to :ACQuire Commands" on page 175
NOTE
Segmented memory is available on the DSOX1000-Series oscilloscope models that have the
SGM license.