Datasheet

Page 20
Find us at www.keysight.com
Built-in voltage/current monitor specifications
Voltage measurement specifications
Range Measurement Resolution Accuracy (% reading + offset)
±200 mV 10 µV ±(0.015 % + 225 μV)
±2 V 100 μV ±(0.02 % + 350 μV)
±20 V 1 mV ±(0.015 % + 5 mV)
±200 V 10 mV ±(0.015 % + 50 mV)
Current measurement specifications
Range Measurement Resolution Accuracy (% reading + offset)
±10 nA 1 pA ±(0.10 % + 50 pA)
±100 nA 10 pA ±(0.06 % + 100 pA)
±1 μA 100 pA ±(0.025 % + 500 pA)
±10 μA 1 nA ±(0.025 % + 1.5 nA)
±100 μA 10 nA ±(0.02 % + 25 nA)
±1 mA 100 nA ±(0.02 % + 200 nA)
±10 mA 1 µA ±(0.02 % + 2.5 μA)
±100 mA 10 µA ±(0.02 % + 20 μA)
±1 A 100 μA ±(0.03 % + 1.5 mA)
±1.5 A 100 μA ±(0.05 % + 3.5 mA)
±3 A 1 mA ±(0.4 % + 7 mA)
±10 A
1
1 mA ±(0.4 % + 25 mA)
2
1. 10 A range is available only for pulse mode, not available for DC mode.
2. Measurement speed: 0.01 PLC
Built-in voltage/current monitor supplemental characteristics
Temperature coefficient(0 to 18 °C and 28 °C to 50 °C) ± (0.1 x Accuracy) /°C
Over range
102 % of measurement range for 1.5 A and 3 A ranges
106 % of measurement range other than 1.5 A and 3 A ranges
Voltage measurement range change overshoot < 250 mV. 100 kΩ load, 20 MHz bandwidth
Current measurement range change overshoot < 250 mV/R load, 20 MHz bandwidth
Derating accuracy for measurement speed less than 1 PLC: Add % of range using the following table for measurement with PLC < 1
Derating accuracy with PLC setting< 1 PLC
Voltage Range Current Range
0.2 V 2 V to 200 V 10 nA 100 nA 1 μA to 100 mA 1 A to 3 A
0.1 PLC 0.01 % 0.01 % 0.1 % 0.01 % 0.01 % 0.01 %
0.01 PLC 0.05 % 0.02 % 1 % 0.1 % 0.05 % 0.02 %
0.001 PLC 0.5 % 0.2 % 5 % 1 % 0.5 % 0.2 %