Datasheet

Page 19
Find us at www.keysight.com
Low Noise Filter (N1294A-022)
Maximum output range 210 V / 3 A (DC)
Output connector Banana, 2-wire/4-wire
Output / residual resistance 0.3 Ω nominal (2-wire)
Small signal bandwidth 2 kHz nominal (1 A and 3 A ranges), 800 Hz nominal (100 mA range)
Noise
Voltage source 0.1 to 10 Hz Same as voltage specification
10 to 20 MHz 350 μVrms (20 V / 1.5 A range, 50 Ω load)
Current source
0.1 to 10 Hz: Same as current specification
10 to 1 MHz 450 μArms (20 V / 1.5 A range, 0.67 Ω load)
Settling time
Voltage source
1
640 μs (2-wire/4-wire)
Current source
2
1.2 ms (2-wire/4-wire)
1. Time required to reach within 0.1 % of final value at open load condition. Step is 10 % to 90 % range. At 20 V range, 1.5 A limit/1.5 A range
2. Time required to reach within 0.1 % of final value at short condition. Step is 10 % to 90 % range. At 1.5 A range, 20 V limit / 20V range
Load transient recovery time: Time to recover to within the settling band following a load change.
800 mA current change
1
Settling band ± 20 mV
Time 450 µs (2-wire), 650 µs (4-wire)
1. At a load change to change the flowing current from +100 mA to +900 mA, 10 V voltage force, 20 V range, 1.5 A limit
Supported
ranges
Voltage 200 mV to 200 V ranges (210 V max)
Current 10 μA to 3 A ranges (3 A max)
Maximum capacitive load 1 mF
Dimensions
41.5 mm H x 58.2 mm W x 127.5 mm D (When the filter is inserted to the output connector of B2961A/
B2962A, the depth is 112.5 mm.)
Weight 0.25 kg
Note:
The 10 A pulse range and programmable output resistance capability are not supported by external filters.
For the current output, only DC is supported when using with external filters.
The current measurement data monitored by the built-in voltage/current monitor can be influenced by the charge and discharge current of the capacitance inside of the filters.
Keysight B2900A Family
The B2900A series is a member of B2900 Precision Instrument Family, which provide a variety of precision measurement solutions
with both sourcing and measurement capabilities. The B2900A series of Source Measure Units (SMUs) have 6.5-digit resolution,
enabling 100 nV/10 fA sourcing and measurement. The B2980A series of picoammeter/electrometer have down to 0.01fA and up
to 10 PΩ. Further information on the B2900 Precision Instrument Family is available at http://www.keysight.com/find/b2900a.
B2980A Picoammeter and Electrometer/High R meter
B2900A Source Measure Unit