Datasheet
Page 14
Find us at www.keysight.com
Sweep capabilities
Sweep mode Linear, logarithmic(log) or list
Sweep direction single or double
Type DC, or pulse
Number of steps 1 to 100,000
Min programmable value to create list sweep waveform Minimum 10 μs with 1 μs resolution
Arbitrary Waveform Generation Capabilities
Pre-defined waveforms:
Supported waveforms Sine, Square, Ramp, Triangle, Trapezoid and Exponential
Frequency 1 mHz to 10 kHz
Programmable frequency/timing resolution: 1 μHz nominal (sine), 250 ns(waveforms other than sine)
Frequency accuracy of time base ±50 ppm
Linearity (best-fit) Voltage Source ±0.01%
1
, Current Source ±0.01%
2
THD Voltage Source -90 dB
3
, Current Source -90 dB
4
1. ±1 V, ±10 V, ±200 V, open load
2. ±10 mA 1 kΩ load, ±100 mA 100 Ω load
3. ±1 V, ±10 V, ±180 V, open load
4. ±1 uA 100 kΩ load, ±10 uA 100 kΩ load, ±10 mA 100 Ω load
User-defined waveforms:
Waveform length 1 to 100,000 points
Sample rate 0.001 to 100,000 Sa/s, 250 ns resolution
Storage
Non-volatile memory and USB memory are both available. Non-volatile memory can store one
waveform with a length of up to 2500 points. USB memory can store waveforms of up to 100k points.
Output characteristics
Temperature coefficient(0 to 18 °C and 28 °C to 50 °C) ± (0.1 x Accuracy) /°C
Max output power and source/sink limits 31.8 W
±6 V@ ±3.03 A, ±21 V@ ±1.515 A, ±210 V@ 105 mA, four quadrant source or sink operation
Output location Channel 1 at front, and channel 2 at rear
Output connectors Banana jack. Triaxial connections are recommended for sourcing less than 1 nA. A banana jack to
triaxial adapter is available for low current source.
Low terminal connection Chassis grounded or floating
Sensing Modes 2-wire or 4-wire (Remote-sensing) connections
Maximum load Normal mode: 0.01 μF
High capacitance mode: 50 μF
DC floating voltage Max ±250 V DC between low force and chassis ground
Guard offset voltage (V source) < 4 mV
Remote sense operation range Max voltage between High Force and High Sense = 3 V
Max voltage between Low Force and Low Sense = 3 V
Common mode isolation > 1 GΩ, < 4500 pF
Max sense lead resistance 1 kΩ for rated accuracy
Sense input impedance > 10 GΩ
Current limit/compliance Accuracy is same as current source. Min value is 1 % of range, or 1 nA in 10 nA range.
Voltage limit/compliance Accuracy is same as voltage source. Min value is 1 % of range, or 20 mV in 200 mV range
Over range
101 % of source range for 1.5 A and 3 A ranges. 105 % of source range other than 1.5 A and 3 A
ranges. No over range for 200 V range with current exceeding 105 mA pulse only condition.
Over temperature protection Output turns off then resets at over temperature sensed internally