Datasheet
Front panel operation
Front panel interface 4.3” TFT color display (480x272, with LED backlight) with keypads and rotary knob
View mode Single view, Dual view, Graph view and Roll view
Hardkeys Single Trigger and Auto Trigger control, 10-key, Rotary Knob and Cursors, Channel on/off, View, Cancel/Local
Softkeys Function, System and Input Assist Keys
Indicators Channel (measurement) status, System status
Source/measurement capabilities
Sweep measurement
Number of steps 1 to 100,000
1
Sweep mode Linear, logarithmic (log) or list
Sweep direction Single or double
Type DC, or pulse
Min programmable value to create list
sweep waveform
B2901A/B2902A: Min 20 μs with 1 μs resolution
B2911A/B2912A: Min 10 μs with 1 μs resolution
1. Valid for the models with S/N below or later, firmware rev. 3.1 or later ,and software entitlement certificate are required.
B2901A MY51140470
B2902A MY51140683
B2911A M Y51140213
B2192A MY51140498
“1 to 2,500” for the models with other S/N.
Digitizing/sampling measurement
Min trigger interval (Max speed of
measurement)
B2901A/B2902A: 20 μs (50,000 points/s)
B2911A/B2912A: 10 μs (100,000 points/s)
Data buffers
Max buffer size 100,000 points / channel
260.5 mm
212.8 mm
450 mm
87.7 mm
480 mm
104 mm
20 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet