Datasheet

08 | Keysight | AC6800 Series Basic AC Power Sources - Data Sheet
www.keysight.com/find/AC6800
Supplemental Characteristics
Model AC6801A AC6802A AC6803A AC6804A
Input rating
Nominal input rating 100 to 120 Vrms/200 to 240 Vrms, 50 Hz or 60 Hz, single-phase
Input voltage range 90 to 132 Vrms/180 to 264 Vrms (auto detected when the power is turned on)
Input frequency range 47 Hz to 63 Hz
Apparent power 800 VA or less 1600 VA or less 3200 VA or less 6400 VA or less
Power factor
1
0.9 (typical)
Input current 8 A/4 A or less
(@100 V/200 V)
16 A/8 A or less
(@100 V/200 V)
32 A/16 A or less
(@100 V/200 V)
64 A/32 A or less
(@100 V/200 V)
6.7 A/3.5 A or less
(@120 V/230 V)
13.4 A/7.0 A or less
(@120 V/230 V)
26.8 A/14.0 A or less
(@120 V/230 V)
53.6 A/28.0 A or less
(@120 V/230 V)
Output ratings for AC mode (135 V/270 V range)
Programmable voltage range 0 to 137.5 Vrms/0 to 275 Vrms
Voltage setting resolution 0.1 V
Frequency setting resolution 0.1 Hz
Voltage THD
2
0.5% or less
Efficiency
3
70% or greater
Maximum repetitive peak current
4
15 A/7.5 A 30 A/15 A 60 A/30 A 120 A/60 A
Output ratings for DC mode (135 V/270 V range)
Programmable voltage range –194.5 to 194.5 V/–389 to 389 V
Voltage setting resolution 0.1 V
Maximum repetitive peak current
4
12 A/6 A 24 A/12 A 48 A/24 A 96 A/48 A
Ripple noise
5
0.7 Vrms/1.4 Vrms (typical)
Output voltage stability
Temperature coefficient
6
100 ppm/°C (typical)
Output voltage response time
7
150 us (typical)
Line regulation
8
±0.15% of full scale
Measurements (135 V/270 V range)
Voltage measurement resolution 0.1 V
Current measurement resolution 0.01 A
Peak current measurement accuracy
9
±(2% of reading +
0.1 A/0.05 A) (typical)
±(2% of reading +
0.2 A/0.1 A) (typical)
±(2% of reading +
0.4 A/0.2 A) (typical)
±(2% of reading +
0.8 A/0.4 A) (typical)
Power measurement resolution 0.1 W, 1 W (for 1000 W or more)
1. For an output voltage of 100 V/200 V (135 V/270 V range), maximum rms current, and a load power factor of 1
2. At an output voltage of 50 to 135 V/100 to 270 V, a load power factor of 1, and in AC mode
3. For AC mode, at an output voltage of 100 V/200 V, maximum rms current, load power factor of 1, and an output frequency of 40 to 500Hz
4. For a periodic, pulsed current waveform with the total rms current under the specified maximum rms current
5. For 5 Hz to 1 MHz components measured at the output terminals on the rear panel
6. For an output voltage of 100 V/200 V, an output current of 0 A, and within the operating temperature range
7. For DC mode, an output voltage of 100 V/200 V, a load power factor of 1, and settling within a band of ±1% about nominal, under a step change in output
current from 0A to the maximum rms current (or its reverse)
8. For line voltage and frequency changes within the rated range
9. For a waveform of crest factor 3 or less, an output current in the range of 5 to 100% of the Maximum repetitive peak current, and 23 ±5 °C