Chapter 3 23
Operation
Fixture Compensation
2. Press OPEN key to start the OPEN compensation data measurement. When the OPEN
compensation data measurement is completed, the softkey label OPEN
on OFF (if it
is off) changes to OPEN
ON off.
Performing Short Compensation
1. Put the probe adapter into the SHORTed state as shown in Figure 3-6. Use an
appropriate device for shorting since no short device is supplied with the BNC
adapter.
NOTE For short compensation, we recommend using a short device with gold-plated
surfaces, which provide stable contact resistance.
Figure 3-6 Performing Short Compensation
2. Press SHORT key to start the SHORT compensation data measurement.When the
SHORT compensation data measurement is completed, the softkey label
SHORT
on OFF (if it is off) changes to SHORT ON off.