Datasheet

20 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet
General Characteristics
(for all models except where noted)
Line power
Power supply: 100/120 (127)/ 220 (230)/240 ACV ± 10%, CAT II
Power line frequency: 50/60/400 Hz ± 10%
Power consumption: 25 VA
Environment
Operating environment: Full accuracy for 0 to 55 °C
Full accuracy to 80% R.H. at 40 °C non–
condensing
Operating altitude: Up to 3,000 m
Storage temperature: 40 to 70 °C
Mechanical
Rack dimensions: (W x H x D): 212.8 mm x 88.3 mm x 272.3 mm
Bench dimensions: (W x H x D): 261.2 mm x 103.8 mm x 303.2 mm
Weight: 34460A: 3.68 kg (8.1 lb)
34461/65/70A: 3.76 kg (8.3 lb)
Regulatory
Safety
EN 61010-1:2010 (3rd Edition)
ANSI/ISA-61010-1 (82.02.01) Third Edition
ANSI/UL 61010-1 Third Edition
CAN/CSA-C22.2 No. 61010-1 Third Edition
EN 61010-2-030:2010 (1st Edition)
ANSI/ISA-61010-2-030 (82.02.03) First Edition
ANSI/UL 61010-2-030 First Edition
CAN/CSA-C22.2 No. 61010-2-030 First Edition
Refer to Declaration of Conformity for current
revisions
Measurement Category II to 300 V
Other non MAINS circuits to 1,000 Vpk
Pollution Degree 2
EMC IEC 61326
EN 61326
CISPR
ICES-001
AS/NZS 2064.1
Refer to Declaration of Conformity for current
revisions
Acoustic noise (nominal) 45 dBA
Triggering conditions
External input Low–power TTL compatible input programmable
edge triggered
Delay: < 1 µs
Jitter: < 1 µs
Minimum pulse width: 1 µs
Maximum rate: Up to 1 kHz (34461A), up to 300 Hz (34460A)
Voltmeter complete
output
3.3 V logic output
Polarity: Programmable edge pulse
Pulse width: Approximately 2 µs
Computer interfaces
LXI (rev 1.4) 10/100Base-T Ethernet (Sockets, VXI-11
protocol, Web user interface) (Optional on
34460A)
USB USB 2.0 (USB-TMC488 & MTP protocol)
GPIB Optional GPIB IEEE-488
Language SCPI-1999, IEEE-488.2, 34401A compatible
Front-panel USB host port
Supports USB 2.0 high-speed mass storage (MSC) class devices
Capability: import/export instrument conguration les, save volatile
readings and screen captures
Triggering and memory
Samples per trigger 1 to 1,000,000
Trigger delay 0 to 3600 sec (~1 µs step size)
External trigger delay < 10 µs
External trigger jitter < 1 µs (DC xed range)
Volatile reading memory 10,000 (34461A), 1,000 (34460A)
Probe hold
Capture and navigate stable list of readings
Internal ash le system
80 MB total capacity
Save reading memory to non-volatile memory in CSV format
Store and recall user-dened states, power-off state,
1
and preference les
Save screen captures in BMP or PNG formats
1. Power-off state only when power-down is initiated via front-panel power
switch.
Math functions
Per function null, min/max/avg/Sdev, dB, dBm, span, count, limit test,
histogram
Display
4.3” color TFT WQVGA (480x272) with LED backlight
Supports: basic number, bar meter, trend chart (34461A only), histogram
views. User-dened power-on message, display label, and selectable screen
colors Integrated, context-sensitive system help through press-and-hold
buttons
Real-time clock/calendar
Set and read, year, month, day, hour, minute, seconds (Note: seconds not set
-
table). Battery CR-2032 coin-type, replaceable, > 10-year life (typ)
Software available
IO Libraries: www.keysight.com/find/IOLibraries
BenchVue: www.keysight.com/find/benchvue
System speeds (nom)
Benchmark GPIB USB 2.0 VXI-11 Sockets
Function change
1
50/s 50/s 50/s 50/s
Range change
2
100/s 100/s 100/s 100/s
1. Rate to change from 2-wire resistance to any other function
2. Rate to change from one range to the next higher range, ≤ 10 V, ≤ 10 M