Datasheet
19 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet
System speeds (nom)
DC voltage, DC current, resistance
1, 2
34460A 34461A 34465A/34470A
Autorange time
3
< 30 ms < 30 ms <5 ms
Maximum internal trigger rate 300/s 1000/s 5,000/s
Maximum external trigger rate 300/s 1000/s 5,000/s
ASCII readings to bus 300/s 1000/s 40,000/s (GPIB 8,000/s)
Single reading transaction rate
4
50/s 150/s 250/s
AC voltage, AC current
5
Autorange time
3
10/s 10/s < 5 ms
Maximum internal trigger rate 50/s 50/s 250/s
Maximum external trigger rate 50/s 50/s 250/s
ASCII readings to bus 50/s 50/s 250/s
Single reading transaction rate
4
50/s
50/s
5
200/s
Frequency, period
6
Autorange time
3
10/s 10/s < 5 ms
Maximum internal trigger rate 80/s 80/s 800/s
Maximum external trigger rate 80/s 80/s 800/s
ASCII readings to bus 80/s 80/s 900/s
Single reading transaction rate
4
50/s 50/s 200/s
1. 0.02 NPLC, delay 0, autozero off, math off, and display off.
2. These rates apply to all I/O interfaces.
3. Time to automatically change one range and be ready for new measurement, ≤ 10 V, ≤ 10 MΩ.
4. Includes measurement and IO time (assumes connection via SOCKETS. VXI-11 connections may be slower).
5. Fast AC lter, delay 0, math off, and display off.
6. 10 ms aperture, fast AC lter, delay 0, math off, and display off.
34460A DMM rear panel with GPIB option installed.
34461/65/70A DMM rear panel with GPIB option installed.