Datasheet

14 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet
Specications 34470A
34470A accuracy specifications: ± (% of reading + % of range)
1
DC voltage and resistance. Automatic calibration (ACAL) capable.
Range
2
24 hours
3
T
ACAL
± 1 °C
90 days
T
ACAL
± 5 °C
1 year
T
ACAL
± 5 °C
2 years
T
ACAL
± 5 °C
Non ACAL
11
Temperature
coefcientC
With ACAL
12
Temperature
coefcientC
DC voltage
100 mV 0.0030 + 0.0030 0.0040 + 0.0035 0.0040 + 0.0035 0.0045 + 0.0035 0.0005 + 0.0005 0.0001 + 0.0005
1 V 0.0010 + 0.0004 0.0015 + 0.0004 0.0020 + 0.0004 0.0025 + 0.0004 0.0005 + 0.0001 0.0001 + 0.0001
10 V 0.0008 + 0.0002 0.0013 + 0.0002 0.0016 + 0.0002 0.0020 + 0.0002 0.0005 + 0.0001 0.0001 + 0.0001
100 V 0.0020 + 0.0006 0.0032 + 0.0006 0.0038 + 0.0006 0.0040 + 0.0006 0.0005 + 0.0001 0.0001 + 0.0001
1000 V
14
0.0020 + 0.0006 0.0032 + 0.0006 0.0038 + 0.0006 0.0040 + 0.0006 0.0005 + 0.0001 0.0001 + 0.0001
Resistance
9
100 Ω 0.0030 + 0.0030 0.0050 + 0.0040 0.0060 + 0.0040 0.0070 + 0.0040 0.0006 + 0.0005 0.0002 + 0.0005
1 K 0.0020 + 0.0005 0.0030 + 0.0005 0.0040 + 0.0005 0.0050 + 0.0005 0.0006 + 0.0001 0.0002 + 0.0001
10 K 0.0020 + 0.0005 0.0030 + 0.0005 0.0040 + 0.0005 0.0050 + 0.0005 0.0006 + 0.0001 0.0002 + 0.0001
100 K 0.0020 + 0.0005 0.0030 + 0.0005 0.0040 + 0.0005 0.0050 + 0.0005 0.0006 + 0.0001 0.0002 + 0.0001
1 MΩ 0.0020 + 0.0005 0.0060 + 0.0005 0.0070 + 0.0005 0.0080 + 0.0005 0.0010 + 0.0002 0.0002 + 0.0002
10 MΩ 0.010 + 0.001 0.020 + 0.001 0.025 + 0.001 0.030 + 0.001 0.0030 + 0.0004 0.0030 + 0.0004
100 MΩ 0.100 + 0.001 0.200 + 0.001 0.300 + 0.001 0.400 + 0.001 0.1000 + 0.0001 0.0100 + 0.0001
1000 M 2.000 + 0.001 2.000 + 0.001 3.000 + 0.001 4.000 + 0.001 1.0000 + 0.0001 0.1000 + 0.0001
Range
2
24 hours
3
T
CAL
± 1 °C
90 days
T
CAL
± 5 °C
1 year
T
CAL
± 5 °C
2 years
T
CAL
± 5 °C
Temperature
coefcientC
6
DC current
1 µA
4
0.007 + 0.005 0.030 + 0.005 0.050 + 0.005 0.060 + 0.005 0.0020 + 0.0010
10 µA
4
0.007 + 0.002 0.030 + 0.002 0.050 + 0.002 0.060 + 0.002 0.0015 + 0.0006
100 µA
4
0.007 + 0.001 0.030 + 0.001 0.050 + 0.001 0.060 + 0.001 0.0015 + 0.0004
1 mA 0.007 + 0.003 0.030 + 0.005 0.050 + 0.005 0.060 + 0.005 0.0015 + 0.0005
10 mA 0.007 + 0.020 0.030 + 0.020 0.050 + 0.020 0.060 + 0.020 0.0020 + 0.0020
100 mA 0.010 + 0.004 0.030 + 0.005 0.050 + 0.005 0.060 + 0.005 0.0020 + 0.0005
1 A 0.050 + 0.006 0.070 + 0.010 0.080 + 0.010 0.100 + 0.010 0.0050 + 0.0010
3 A 0.180 + 0.020 0.200 + 0.020 0.200 + 0.020 0.230 + 0.020 0.0050 + 0.0020
10 A
5
0.050 + 0.010 0.120 + 0.010 0.120 + 0.010 0.150 + 0.010 0.0050 + 0.0010
Continuity
1 K 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 0.012 + 0.020 0.0010 + 0.0020
Diode test
11
5 V 0.002 + 0.010 0.008 + 0.010 0.010 + 0.010 0.012 + 0.020 0.0010 + 0.0010
DC:DC ratio
4, 12
(normalized input accuracy) + (normalized reference accuracy)