Datasheet

DC specification
Table 1. DC accuracy ± (% of reading + % of range)
Function Range
1
Test current or
Burden Voltage
Input
Impedance
1 Year
23 °C ± 5 °C
Temperature Coefficient / °C
0 °C - 18 °C
28 °C - 55 °C
DC voltage 100.000 mV 10 MΩ or > 10 GΩ 0.018 + 0.008 0.0020 + 0.0008
1.00000 V 10 MΩ or > 10 GΩ 0.015 + 0.005 0.0015 + 0.0008
10.0000 V 10 MΩ 0.015 + 0.005 0.0020 + 0.0008
100.000 V 10 MΩ 0.015 + 0.005 0.0020 + 0.0008
1000.00 V 10 MΩ 0.015 + 0.005 0.0020 + 0.0008
Resistance
(2-wire)
2
Resistance (4-wire)
2
100.000 Ω 1 mA 0.050 + 0.008 0.0060 + 0.0008
1.00000 500 μA 0.050 + 0.008 0.0060 + 0.0005
10.0000 kΩ 100 μA 0.050 + 0.005 0.0060 + 0.0005
100.000 kΩ 10 μA 0.050 + 0.005 0.0060 + 0.0005
1.00000 MΩ 1 μA 0.060 + 0.005 0.0060 + 0.0005
10.0000 MΩ 100 nA 0.250 + 0.005 0.0250 + 0.0005
100.000 MΩ 100 nA/10 MΩ 2.000 + 0.005 0.3000 + 0.0005
DC current 100.000 μA < 0.02 V 0.05 + 0.015 0.007 + 0.0015
1.00000 mA < 0.2 V 0.05 + 0.007 0.007 + 0.0010
10.0000 mA < 0.02 V 0.05 + 0.015 0.008 + 0.0015
100.000 mA < 0.2 V 0.05 + 0.007 0.008 + 0.0010
1.00000 A < 0.2 V 0.10 + 0.015 0.012 + 0.0015
10.0000 A < 0.6 V 0.25 + 0.007 0.015 + 0.0010
Continuity
3
1000 Ω 0.5 mA 0.1 + 0.1 0.005 + 0.005
Diode test
4
1.0000 V 0.5 mA 0.05 + 0.03 0.005 + 0.005
Notes on specifications:
1. 20% over range on all ranges except 1000 VDC and 10 A range.
2. Specifications are for 4-wire Ω or 2-wire Ω using NULL function. If without NULL function, add 0.2 Ω additional error.
3. Continuity thresholds is fixed at leas than 10 Ω. Available in fast mode only.
4. Specifications are for the voltage measured at the input terminals only. Available in fast mode only.
Specifications are for 90 minutes warm-up time, slow mode, NULL function enabled, and calibration temperature within 18 °C - 28 °C
(unless stated otherwise)
Specifications
Specifications are for 90 minutes warm-up time, slow
mode, and calibration temperature within 18 °C - 28 °C
Page 4
Find us at www.keysight.com