Datasheet
General characteristics
Table 8. General Characteristics
Specification measurement Specification
Power supply – 100 V/120 V(127 V)/220 V(230 V)/240 V ± 10%
– AC line frequency 45 Hz - 66 Hz and (360 Hz - 440 Hz, 100/120 V operation)
– Automatically sensed at power -on
Power consumption 45 VA maximum, < 11 W average
Operating environment Full accuracy to 80% RH for 0 °C to 30 °C (non-condensing)
Full accuracy to 40% RH for 30 °C to 55 °C (non-condensing)
Altitude up to 3000 meters
Operating temperature Full accuracy for 0 °C to 55 °C
Storage compliance –40 °C to 70 °C
Safety compliance – IEC 61010-1:2001/EN 61010-1:2001 (2nd Edition)
– Canada: CAN/CSA-C22.2 No. 61010 -1 -04
– USA: ANSI/UL std No. 61010-1:2004
Measurement category CAT II, 300 V: CAT I 1000 Vdc, 750 Vac rms, 2500 Vpk transient over voltages, Pollution degree 2
EMC compliance Certified to IEC61326-1: 2005/EN61326-1:2006
CISPR 11:2003 / EN 55011:2007 Group 1 Class A
Canada: ICES/NMB-001:Issue 4, June 2006
Australia/New Zealand: AS/NZS CISPR 11:2004
Shock and vibration Tested to IEC/EN 60086-2
Dimension (HxWxD) Rack: 88.5 mm x 212.6 mm x 272.3 mm
Bench: 103.8 mm x 261.1 mm x 303.2 mm
Weight 3.75 kg (8.27 lb.)
Warm up time 90 minutes
303.2 mm
103.8 mm
261.1 mm
Standard shipped accessories
– Test lead set
– Power cord
– USB interface cable
– Quick Start Guide
– Product Reference CD
– Keysight IO Library Suite CD-ROM
Upgradable options
– 3445GPBU - GPIB connectivity
upgrade
– 3445MEMU - 50,000 memory
points upgrade for data logging
Dimensions
Page 10
Find us at www.keysight.com