Datasheet
14
Math functions
Per function null, min/max/avg/Sdev, dB, dBm, span, count, limit test,
histogram
Display
4.3” color TFT WQVGA (480x272) with LED backlight
Supports: basic number, bar meter, trend chart (34461A only), histogram views
User-defined power-on message, display label, and selectable screen colors
Integrated, context-sensitive system help through press-and-hold buttons
Real-time clock/calendar
Set and read, year, month, day, hour, minute, seconds (Note: seconds not
settable)
Battery CR-2032 coin-type, replaceable, >10-year life (typ)
Software available
IO Libraries: www.Keysight.com/find/IOLibraries DMM Connectivity Utility
software:
www.Keysight.com/find/DMMutilitysoftwareSystem
34460A/34461A
General Characteristics 34460A / 34461A
General characteristics
Line power
Power supply: 100/120 (127)/ 220 (230)/240 VAC ± 10%, CAT II
Power line frequency: 50/60/400 Hz ± 10%
Power consumption: 25 VA
Environment
Operating environ-
ment:
Full accuracy for 0 to 55 °C
Full accuracy to 80% R.H. at 40 °C non–condensing
Operating altitude: Up to 3000 m
Storage temperature: -40 to 70 °C
Mechanical
Rack dimensions: (W x H x D): 212.8 mm x 88.3 mm x 272.3 mm
Bench dimensions: (W x H x D): 261.2 mm x 103.8 mm x 303.2 mm
Weight: 34460A: 3.68 kg (8.1 lb)
34461A: 3.76 kg (8.3 lb)
Triggering and memory
Samples per trigger 1 to 1,000,000
Trigger delay 0 to 3600 sec (~1 s step size)
External trigger delay <10 s
External trigger jitter <1 s (DC fixed range)
Volatile reading
memory
10,000 (34461A), 1,000 (34460A)
Probe hold
Capture and navigate stable list of readings
Internal flash file system
80 MB total capacity
Save reading memory to non-volatile memory in CSV format
Store and recall user-defined states, power-off state,
1
and preference files
Save screen captures in BMP or PNG formats
System speeds (averages)
Benchmark GPIB USB 2.0 VXI-11 Sockets
Function change
1
50/s 50/s 50/s 50/s
Range change
2
100/s 100/s 100/s 100/s
1. Rate to change from 2-wire resistance to any other function
2. Rate to change from one range to the next higher range, ≤ 10 V, ≤ 10 MΩ
1. Power-off state only when power-down is initiated via front-panel power switch.
Regulatory
Safety EN 61010-1:2010 (3rd Edition)
ANSI/ISA-61010-1 (82.02.01) Third Edition
ANSI/UL 61010-1 Third Edition
CAN/CSA-C22.2 No. 61010-1 Third Edition
EN 61010-2-030:2010 (1st Edition)
ANSI/ISA-61010-2-030 (82.02.03) First Edition
ANSI/UL 61010-2-030 First Edition
CAN/CSA-C22.2 No. 61010-2-030 First Edition
Refer to Declaration of Conformity for current revisions
Measurement Category II to 300 V
Other non MAINS circuits to 1000 Vpk
Pollution Degree 2
EMC IEC 61326
EN 61326
CISPR
ICES-001
AS/NZS 2064.1
Refer to Declaration of Conformity for current revisions
Acoustic noise (nominal) 45 dBA
Acoustic noise
(nominal)
45 dBA
Triggering conditions
External input Low–power TTL compatible input programmable edge
triggered
Delay: <1 s
Jitter: <1 s
Minimum pulse
width:
1 s
Maximum rate: Up to 1 kHz (34461A), up to 300 Hz (34460A)
Voltmeter complete
output
3.3 V logic output
Polarity: Programmable edge pulse
Pulse width: Approximately 2 s
Computer interfaces
LXI (rev 1.4) 10/100Base-T Ethernet (Sockets, VXI-11 protocol,
Web user interface) (Optional on 34460A)
USB USB 2.0 (USB-TMC488 & MTP protocol)
GPIB Optional GPIB IEEE-488
Language SCPI-1999, IEEE-488.2, 34401A compatible
Front-panel USB host port
Supports USB 2.0 high-speed mass storage (MSC) class devices
Capability: import/export instrument configuration files, save volatile readings
and screen captures