Datasheet
08 | Keysight | 53131A/132A/181A Counters - Data Sheet
Measurement Specications
Frequency (53131A, 53132A, 53181A)
Channel 1 and 2 (53131A, 53132A)
Channel 1 (53181A)
Range 0.1 Hz to 225 MHz
Channel 3 (53131A, 53132A)
Channel 2 (53181A)
Option 015 (53181 A only) 100 MHz to 1.5 GHz
Option 030 100 MHz to 3 GHz
Option 050 200 MHz to 5 GHz
Option 124 200 MHz to 12.4 GHz
(Period 2 or 3 selectable via GPIB only)
Period (53131A, 53132A, 53181A)
Channel 1 and 2 (53131A, 53132A)
Channel 1 (53181A)
Range 4.44 ns to 10 s
Channel 3 (53131A, 53132A) Option 015 (53181A only) 0.66 to 10 ns
Channel 2 (53181A) Option 030 0.33 to 10 ns
Option 050 0.2 to 5 ns
Option 124 80 ps to 5 ns
Frequency ratio (53131A, 53132A, 53181A)
Measurement is specied over the full signal
range of each input.
Results range 10
–10
to 10
11
“Auto” gate time 100 ms
Time interval (53131A, 53132A)
Measurement is specied over the full signal
ranges
6
of Channels 1 and 2
Results range –1 ns to 10
5
s
LSD 500 ps (53131A)/150 ps (53132A)
Phase (53131A, 53132A)
Measurement is specied over the full signal
range of Channels 1 and 2
Results range –180° to +360°
Duty cycle (53131A, 53132A)
Measurement is specied over the full signal
range of Channel 1. However, both the positive
and negative pulse widths must be greater than
4 ns.
Results range 0 to 1 (e.g. 50% duty cycle would be displayed
as .5)
Rise/fall time (53131A, 53132A)
Measurement is specied over the full signal
ranges of Channel 1. The interval between the
end of one edge and start of a similar edge must
be greater than 4 ns.
Edge selection Positive or negative
Trigger Default setting is auto trigger at 10 and 90%
Results range 5 ns to 10
5
s
LSD 500 ps (53131A)/150 ps (53132A)
Pulse width (53131A, 53132A)
Measurement is specied over the full signal
range of Channel 1. The width of the opposing
pulse must be greater than 4 ns.
Pulse selection Positive or negative
Trigger Default setting is auto trigger at 50%
Results range 5 ns to 10
5
s
LSD 500 ps (53131A)/150 ps (53132A)
Totalize (53131A, 53132A)
Measurement is specied over the full signal
range of Channel 1.
Results range 0 to 10
15
Resolution ± 1 count
Peak volts (53131A, 53132A, 53181A)
Measurement is specied on Channels 1 and 2
for DC signals; or for ac signals of frequencies
between 100 Hz and 30 MHz with peak-to-peak
amplitude greater than 100 mV.
Results range –5.1 V to +5.1 V
Resolution 10 mV
Peak volts systematic uncertainty (53131A, 53132A, 53181A)
Use of the input attenuator multiplies all voltage
specications (input range, results range,
resolution and systematic uncertainty) by a
nominal factor of 10.
For AC signals 25 mV + 10% of V
For DC signals 25 mV + 2% of V