Datasheet
7
[1] Typical at full-scale reading and voltage across the input terminals. 
[2] In 10 A range, >10 to 20 ADC is readable for 20 seconds maximum with audio warning.
[1] Typical at full-scale reading and voltage across the input terminals. 
[2] Relative (REL) operation is used.
DC current
DCI resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM 
U3402A 5.5-digit DMM
Range Resolution  Maximum reading  Accuracy (1 year; 23 ºC ± 5 ºC)  Burden voltage 
[1]
 and shunt resistor 
500.00 µA 10 nA  510.00  0.05% + 5 
<0.06 V/100 W 
5.0000 mA  100 nA  5.1000  0.05% + 4
<0.6 V/100 W
50.000 mA 1 µA 51.000  0.05% + 4
<0.08 V/1 W
500.00 mA  10 µA 510.00 0.05% + 4
<0.8 V/1 W 
5.0000 A 100 µA 5.1000 0.25% + 5
<0.3 V/0.01 W
10.000 A 1 mA 20.000 
[2]
0.25% + 5
<0.6 V/0.01 W
Rate Range  Resolution Maximum reading Accuracy 
(1 year; 23 ºC ± 5 ºC)
Burden voltage
 [1]
and shunt resistor 
Slow  12.0000 mA 
120.000 mA 
1200.00 mA 
12.0000 A 
0.1 µA 
1 µA 
10 µA
100 µA 
11.9999 
119.999
1199.99
11.9999 
0.05% + 15 
[2]
0.05% + 5 
0.2% + 5 
0.2% + 5
<0.15 V/10 W 
<1.5 V/10 W 
<0.3 V/0.1 W  
<0.6 V/0.01 W 
Medium 40.000 mA
120.00 mA
1200.0 mA 
12.000 A 
1 µA 
10 µA 
100 µA 
1 mA
39.999 
119.99 
1199.9 
11.999
0.1% + 6 
0.1% + 3 
0.2% + 3 
0.2% + 3
<0.5 V/10 W 
<1.5 V/10 W 
<0.3 V/0.1 W 
<0.6 V/0.01 W 
Fast 40.00 mA 
120.0 mA 
1200 mA 
12.00 A 
10 µA 
100 µA 
1 mA 
10 mA
39.99 
119.9
1199 
11.99
0.1% + 2 
0.1% + 2 
0.2% + 2 
0.2% + 2 
<0.5 V/10 W
<1.5 V/10 W 
<0.3 V/0.1 W
<0.6 V/0.01 W 