Datasheet
11
[1] Specified accuracy at input >5% of full scale.
Frequency
Frequency resolution, full-scale reading and accuracy [± (% of reading + count)]
Diode test/continuity
U3401A 4.5-digit DMM
U3401A 4.5-digit DMM
U3402A 5.5-digit DMM
U3402A 5.5-digit DMM
Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC)
2.3000 V 100 µV 2.3000 V 0.05% + 5
Rate Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC)
Slow 1.19999 V 10.0000 µV 0.012% + 5
Medium 2.4999 V 100.00 µV 0.012% + 5
Fast 2.499 V 1.000 mV 0.012% + 2
Range Input sensitivity for voltage measurement (sine-wave)
5 Hz to 100 kHz 100 kHz to 500 kHz
500 mV 35 mV RMS 200 mV RMS
5 V 0.25 V RMS 0.5 V RMS
50 V 2.5 V RMS 5 V RMS
500 V 25 V RMS Not specified
750 V 50 V RMS Not specified
Range Measurement range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC)
500.00 Hz 5 Hz to 500 Hz 0.01 Hz 510.00 0.01% + 5
5.0000 kHz
500 Hz to 5 kHz
0.1 Hz 5.1000 0.01% + 3
50.000 kHz 5 kHz to 50 kHz 1 Hz 51.000 0.01% + 3
500.00 kHz 50 kHz to 500 kHz 10 Hz 999.99
0.01% + 3
Range Measurement range Resolution Maximum reading Accuracy
(1 year; 23 ºC ± 5 ºC)
[1]
Input sensitivity
(sine wave)
1200.00 Hz 5 Hz to 1200 Hz 10 mHz 1199.99 0.005% + 3 40 mV RMS
12.0000 kHz
10 Hz to 12 kHz
100 mHz 11.9999 0.005% + 2 40 mV RMS
120.000 kHz 100 Hz to 120 kHz 1 Hz 119.999 0.005% + 2 40 mV RMS
1.0000 MHz 1 kHz to 1 MHz 10 Hz 1.1999
0.005% + 2
0.5 V RMS