Datasheet
70
Waveform characteristics (continued)
Arbitrary
Frequency range 1 μHz to 200 kHz (1 μHz resolution)
Waveform memory depth 64 kSa
[1]
Amplitude resolution 14 bits/sample (including sign)
Sampling rate 50 MSa/s
Minimum rise/fall time 36 ns (Typical)
Linearity < 0.2% of peak output
Settling Time < 250 ns to 0.5% of fi nal value
Jitter (RMS) 10 ns + 30 ppm
Common characteristics
Amplitude
Range
40 mVpp to 5 Vpp (Into 50 Ω load)
80 mVpp to 10 Vpp (Into open circuit)
Accuracy
[2]
(across 50 Ω load at 1 kHz)
±1% of setting ± 5 mV (±10 mV @ Hi-Z)
Units Vpp, Vrms, dBm
Resolution 4 digits
DC offset
Range (peak AC + DC)
±2.5 V (Into 50 Ω load)
±5 V (Into open circuit)
Accuracy
[2]
(across 50 Ω load)
±2% of offset setting
±1% of amplitude
±5 mV (±10 mV @Hi-Z)
Amplitude Limit
Amplitude + Offset limit to witthin ±2.5 V range across 50 Ω load or ±5 V across open
circuit
Main output
Impedance 50 Ω load (Typical)
Isolation At least 42 Vpk to earth
Protection Short-circuit protected, overload automatically disables main output
Internal frequency reference
Accuracy
[3]
±8 ppm in 1 year
External frequency reference
Input
• Lock range
• Amplitude level
• Impedance
• Lock time
10 MHz ±170 Hz
500 mVpp to 5 Vpp
50 Ω AC coupled
< 2 s
Output
• Frequency
• Amplitude Level
• Impedance
10 MHz
632 mVpp (Typical)
Return loss 10 dB (Typical) at 10 MHz
Phase Offset
• Range
• Resolution
• Accuracy
+360° to –360°
0.01°
20 ns
[1]
Maximum at 16 k points for Arbitrary waveforms when using bundled software, Agilent Measurement Manager (AMM) and 64 k points when
programmed in compatible application development environments like Agilent VEE, NI LabVIEW, and Microsoft Visual Studio.
[2]
Add 1/10th of output amplitude and offset specifi cation per °C for operation outside the range of 18 °C to 28 °C.
[3]
Add 1 ppm/°C (average) for operation outside the range of 18 °C to 28 °C.