Agilent U2531A USB Simultaneous Sampling DAQ in Power Measurement Unit Monitoring Application Note
Introduction Mobile devices are getting more and more popular. The mobile phone in the market now has become a pocket media center. It contains at least a digital still camera, video conference functionality, an MP3 player with simulated surround sound, and a color Internet browser coupled with 3-D gaming capability.
In this application, the program is written in Agilent VEE Pro to control the DAQ devices in acquiring the voltage and current as shown in Figure 3. Figure 3. Windows shown is the VEE test program. The functionality of the PMU is controlled by its firmware. In this particular application, the VEE program is used to control the DAQ devices with regard to the monitoring of the different channels. In addition, a separate routine is written to run through the functionality of the PMU.
The duration of the spike is 0.3 ms. In fact, with the 2 MSa/s sampling rate DAQ, and the recommendation that the sampling rate employed by the DAQ (or the test system) should be five times the sampling rate of the signal, the maximum glitch that can be detected is 2.5 us. In this application, the absolute voltage is actually not as important since the main purpose of the application is to monitor the presence of spikes.
Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. Agilent Direct www.agilent.com/find/agilentdirect Quickly choose and use your test equipment solutions with confidence. Agilent Open www.agilent.com/find/open Agilent Open simplifies the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products.