User manual
Calibration 2
U2300A Series DAQ Service Guide 37
Analog Output
NOTE
Byte ordering must use LSB first.
Step 2: DAQ configuration
Send “SOUR:VOLT:POL Polarity,(@Channel)”,
Send “SOUR:VOLT”RSRC INT,(@Channel)”,
Send “SOUR:VOLT Voltage,(@Channel)”
Step 3: DMM
Measure the output voltage of DAQ
Repeat steps above for all channels and
polarities. The test points are shown as table
Compare the reading with
the test specification
Step 1:
Power on DAQ device and conenct the DMM
to DAQ. Warm up for 30 minutes