User manual
60 U2300A Series DAQ Service Guide
BTest Limits
Analog Input
U2331A:
Offset/
Gain Error (V)
Setting Range (V) Test Point (V) Te st Limit (V) Te st Result
Polarity Min Max Location Value Lower Upper Reading
0.002 BIP –10 10 Center FS 0.00000 -0.00200 0.00200 Offset = DAQ reading – Calibrator source voltage
0.006 BIP –10 10 Positive FS 9.80469 9.79869 9.81069 DAQ reading – Offset
0.006 BIP –10 10 Negative FS –9.80469 –9.81069 –9.79869 DAQ reading – Offset
0.0015 BIP –5 5 Center FS 0.00000 –0.00150 0.00150 Offset = DAQ reading – Calibrator source voltage
0.004 BIP –5 5 Positive FS 4.90234 4.89834 4.90634 DAQ reading – Offset
0.004 BIP –5 5 Negative FS –4.90234 –4.90634 –4.89834 DAQ reading – Offset
0.0015 BIP –2.5 2.5 Center FS 0.00000 –0.00150 0.00150 Offset = DAQ reading – Calibrator source voltage
0.002 BIP –2.5 2.5 Positive FS 2.45117 2.44917 2.45317 DAQ reading – Offset
0.002 BIP –2.5 2.5 Negative FS –2.45117 –2.45317 –2.44917 DAQ reading – Offset
0.001 BIP –1.25 1.25 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.0015 BIP –1.25 1.25 Positive FS 1.22559 1.22409 1.22709 DAQ reading – Offset
0.0015 BIP –1.25 1.25 Negative FS –1.22559 –1.22709 –1.22409 DAQ reading – Offset
0.001 BIP –1 1 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –1 1 Positive FS 0.98047 0.97947 0.98147 DAQ reading – Offset
0.001 BIP –1 1 Negative FS –0.98047 –0.98147 –0.97947 DAQ reading – Offset
0.001 BIP –0.5 0.5 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –0.5 0.5 Positive FS 0.49023 0.48923 0.49123 DAQ reading – Offset
0.001 BIP –0.5 0.5 Negative FS –0.49023 –0.49123 –0.48923 DAQ reading – Offset
0.001 BIP –0.25 0.25 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –0.25 0.25 Positive FS 0.24512 0.24412 0.24612 DAQ reading – Offset
0.001 BIP –0.25 0.25 Negative FS –0.24512 –0.24612 –0.24412 DAQ reading – Offset
0.001 BIP –0.2 0.2 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –0.2 0.2 Positive FS 0.19609 0.19509 0.19709 DAQ reading – Offset
0.001 BIP –0.2 0.2 Negative FS –0.19609 –0.19709 –0.19509 DAQ reading – Offset
0.001 BIP –0.05 0.05 Center FS 0.00000 –0.00100 0.00100 Offset = DAQ reading – Calibrator source voltage
0.001 BIP –0.05 0.05 Positive FS 0.04902 0.04802 0.05002 DAQ reading – Offset
0.001 BIP –0.05 0.05 Negative FS –0.04902 –0.05002 –0.04802 DAQ reading – Offset