User manual
Troubleshooting and Self-Test Procedures 4
U2300A Series DAQ Service Guide 49
Self-Test Procedures
4 Go to Start > All Programs > Agilent IO Libraries Suite > Agilent
Connection Expert to launch the Agilent Connection Expert.
5 Go to Start > All Programs > Agilent T&M Toolkit > Agilent Interactive IO
to launch the Interactive I/O dialog box.
6 Send the SCPI command “*TST?” to the instrument to start
perform the self- test of the instrument.
7 The command will return either "+0" to indicate all tests passes
or "+1" to indicate one or more tests failed.
8 If the command returns "+1" , apply SCPI command
“SYSTem:ERRor?” to enquire the error message.
WARNING
Do not connect any cables and terminal block prior to performing
self-test procedures.
NOTE
Refer to Agilent U2300A Series USB Multifunction Data Acquisition
Programming Guide for SCPI error messages.