User manual

Calibration 2
U2300A Series DAQ Service Guide 29
Option 2: Self-calibration with Agilent Measurement Manager
1 Power on the DAQ device and disconnect all connections
from it. Warm it up for 30 minutes to ensure that it is
operating at stable condition.
2 Connect the DAQ device to the PC with mini- B type USB
cable. Launch the Agilent Measurement Manager and
select the DAQ device you wish to do the self- calibration
process.
3 Go to Tools and select Self Calibration.
4 The Self Calibration Form dialog box will appear as shown
below.
Figure 2-11 Self Calibration Form dialog box in Agilent Measurement
Manager
WARNING
Unplug all cables that are connected to the DAQ device before
performing self-calibration.
Any cables connected to the DAQ device may cause the failure of
the self-calibration process.