User manual
24 U2300A Series DAQ Service Guide
2 Calibration
Analog Output Connection
The equipments required for analog output connection are
the DAQ device, DMM, terminal block, USB mini- B cable and
SCSI cable. Follow the follwing step- by- step instruction for
analog input connection.
1 Connect the DAQ device to a PC with a USB mini- B cable
and connect the DAQ device to a U2901A terminal block
using the U2901A SCSI cable.
2 Connect a wire at each pin you would like to test on the
U2901A terminal block. (Refer to “Connector Pins
Configuration” for the pins assignment)
3 Connect the DAQ device’s analog output channel 1
(AO201) to the DMM input.
4 Connect the DAQ device’s analog output ground
(AO_GND) to DMM GND.
5 Repeat steps 3 and 4 for analog output channel 2
(AO202).
NOTE
• Ensure that the PC has the DAQ device’s driver and the Agilent IO
Libraries 14.2 or higher installed. Note that the Agilent Measurement
Manager software comes with the standard purchase of the U2300A
Series DAQ devices.
• If you do not have the DAQ device’s driver and the Agilent IO Libraries
14.2 or higher installed, refer to the Agilent U2300A Series Data
Acquisition Devices and Agilent Measurement Manager Quick Start
Guide, for more information on the installation.