16 U2300A Series DAQ Service Guide
2 Calibration
Introduction
Prior to checking the performace of the instrument, ensure
that you have all the equipments listed in the following to
perform the verification procedure for analog input and
analog output.
Equipment Checklist
Agilent U2300A Series USB Multifunction DAQ devices
Agilent Digital Multimeter (DMM) 34401A/34410A
Fluke Calibrator 5520A/5720A
Computer with Agilent IO Library Suites installed
U2901A Terminal Block
USB Mini- B Cable
U2901A SCSI Cable
Follow the following steps to ensure the instrument gives
accurate analog input readings and analog outputs.
1 Perform the self- calibration procedure. See
Self- Calibration”.
2 Perform the performance verification procedure. See
Performance Verification Procedure”, “Analog Input
Connection” for information on analog input connection
and “Analog Output Connection” for information on
analog output connection.
3 If the instrument gives accurate readings or accurate
outputs, adjustment procedure is not necessary. If the
instrument does not give accurate readings or accurate
outputs, perform the adjustment procedure. See
Adjustment Procedures”.
4 After perform the adjustment procedure, repeat the
self- calibration procedure and then perform the
verification procedure.
5 If the instrument still does not give accurate readings or
accurate outputs, repeat step 3 and step 4.