Digital multimeter specifications (continued)
Maximum
reading
10,000 counts with automatic polarity indication
Voltage CAT II 1000 V or CAT III 600 V
Function Range Resolution Accuracy Input impedance
(nominal)
Test current
Temperature
3
-50 to 1000 °C 1 mV/°C
-50 to -21 °C 2.5% + 2 °C
7
-20 to 350 °C 0.5% + 2 °C
7
351 to 500 °C 1.75% + 2 °C
7
501 to 1000 °C 2% + 2 °C
7
-58 to 1832 °F 1 mV/°F
-58 to -5.8 °F 2.5% + 3.6 °F
7
-4 to 662 °F 0.5% + 3.6 °F
7
664 to 932 °F 1.75% + 3.6 °F
7
933 to 1832 °F 2% + 3.6 °F
7
Frequency
3
100.00 Hz 0.01 Hz
0.03% + 3
1000.0 Hz 0.1 Hz
10.000 kHz 0.001 kHz
100.00 kHz 0.01 kHz
1000.0 kHz 0.1 kHz
1. Only allowed to measure up to CAT III 600 V if referring to GND.
2. Only allowed for floating voltage.
3. Denotes typical specifications, all others are warranted.
4. The accuracy is specified after the Null function is used to subtract the test lead resistance and thermal effect.
5. RH is specified for < 60%. The temperature coefficient is 0.15 × specified accuracy as > 50 MΩ.
6. The accuracy is based on film capacitors or better and uses the Relative mode for residual values.
7. The accuracy is based on using the Null function to reduce the thermal effect.
8. Denotes characteristics.