Datasheet

9
Electrical Specifications for U1212A (continued)
Capacitance
Range Resolution Accuracy Overload protection
400 µF 0.1 µF 2 % + 4
1000 V R.M.S. for circuits
< 0.3 A of short circuit current
4000 µF 1 µF 3 % + 4
The accuracy is based on film capacitor or better and use Relative mode to zero residual value
DC current
Range Resolution Accuracy Maximum overload
40 A 0.01 A 1.5 % + 15
1000 A R.M.S.400 A 0.1 A 1.5 % + 3
1000 A 1 A 2.0 % + 5
Use Relative mode to zero residual offset
AC current
Range Resolution
Accuracy *N1
Maximum overload45~65 Hz 65 Hz~1 KHz
40 A 0.01 A 2.0 % + 10 3.0 % + 10
1000 A R.M.S.400 A 0.1 A 2.0 % + 5 3.0 % + 5
1000 A 1 A 2.0 % + 5 3.0 % + 5
N1: The maximum verification of current and frequency product is less than 400,000 A x Hz
Current (1 ms peak hold)
Range Resolution Accuracy Maximum overload
40 A 0.01 A 2.0 % + 70
1000 A R.M.S.400 A 0.1 A 2.0 % + 43
1000 A 1 A 2.0 % + 43
Specified accuracy for changes > 1 ms in duration
Temperature test
Thermal type Range Resolution Accuracy
K
–200 ~ –40 °C 0.1 °C 1 % + 3 °C
–40 ~ 1372 °C 0.1 °C 1 % + 1 °C
–328 ~ –40 °F 0.1 °F 1 % + 6 °F
–40 ~ 2502 °F 0.1 °F 1 % + 2 °F
Notes:
1. The accuracy does not include the tolerance of the thermocouple probe, and the meter must be operational for at least one hour
2. Do not allow the temperature sensor to contact a surface that is energized above 33 V R.M.S. or 70 V DC. Such voltages will pose a shock hazard.
3. The temperature calculation is according to the standard of EN/IEC-60548-1 and NIST175