Datasheet
14
Electrical Specifications for U1213A (continued)
Frequency (AC coupling)
Range Resolution Accuracy Minimum frequency
99.99 Hz 0.01 Hz
0.2 % + 3 10 Hz
999.9 Hz 0.1 Hz
9.999 kHz 0.001 kHz
99.99 kHz 0.01 kHz
999.9 kHz 0.1 kHz
• Overload protection: 1000 V; < 20,000,000 V x Hz
Sensitivity
Frequency sensitivity
Range Minimum sensitivity (R.M.S.)
Maximum input for specified accuracy of AC 45 Hz~2 kHz 10 Hz~200 kHz
4 V 0.3 V 0.6 V
40 V 2 V 3 V
400 V 20 V 30 V (< 100 kHz)
1000 V 50 V 50 V (< 10 kHz)
40 A 3 A (< 1 kHz) 3 A (< 1 kHz)
400 A 20 A (< 1 kHz) 20 A (< 1 kHz)
1000 A 50 A (< 1 kHz) 50 A (< 1 kHz)
Duty cycle
Mode Range Accuracy of full scale
AC coupling 0.1 %~99.9 % 0.3 % per kHz + 0.3 %
Notes:
1. The accuracy for duty cycle is based on a 4 V square wave input to the DC 4 V range and maximum frequency up to 2 kHz. The duty cycle range can
be measured within 5 %~95 % as the signal frequency > 20 Hz.