Datasheet
12
Electrical Specifications for U1213A (continued)
Diode check/audible continuity test
Range Resolution Accuracy Test current Open voltage
Diode 0.001 V 0.5 % + 2 Approximately 0.8 mA < +3.1 V
• Overload protection: 1000 V R.M.S. for circuits < 0.3 A of short circuit current
• Built-in buzzer sounds when reading is below 50 mV approximately and single tone for normal forward-biased diode or semiconductor junction as
0.3 V ≤ Reading ≤ 0.8 V
Resistance
Range Resolution Accuracy Test current
400 Ω 0.1 Ω
0.3 % + 3
0.8 mA
4 kΩ 0.001 kΩ 80 µA
40 kΩ 0.01 kΩ 8 µA
400 kΩ 0.1 kΩ 727 nA
4 MΩ 0.001 MΩ 0.6 % + 3 112 nA
40 MΩ 0.01 MΩ 2.0 % + 5 112 nA
Notes:
1. Overload protection: 1000 V R.M.S. for circuits < 0.3 A of short circuit current
2. Maximum open voltage: < +3.1 V
3. Instant continuity: Built-in buzzer sounds when resistance is less than 10.0 Ω
4. The accuracy of 400 Ω and 4 kΩ is specified after Relative function, which is used to substrate the test lead resistance and the thermal effect
Capacitance
Range Resolution Accuracy Overload protection
4 µF 0.001 µF
1 % + 4
1000 V R.M.S. for circuits
< 0.3 A of short circuit current
40 µF 0.01 µF
400 µF 0.1 µF 2 % + 4
4000 µF 1 µF 3 % + 4
• The accuracy is based on film capacitor or better and use Relative mode to zero residual value
DC current
Range Resolution Accuracy Maximum overload
40 A 0.01 A 1.5 % + 15
1000 A R.M.S.400 A 0.1 A 1.5 % + 3
1000 A 1 A 2.0 % + 5
• Use Relative mode to zero residual offset