ELECTRICAL SPECIFICATIONS FOR U1213A
Accuracy is given as ± % of reading +
no. of least signifi cant digits at 23 °C ±
5 °C, with relative humidity Less than
80% R.H.
AC voltage, AC current, AC+DC
voltage, AC+DC current specifi cations
are ac coupled, true R.M.S. and are
valid from 5 % of range to 100 % of
range. The crest factor may be up to
3.0 at full-scale except the 1000 V
and 1000 A ranges where these are
1.5 at full scale. For non-sinusoidal
waveforms, add (2 % reading + 2 %
full scale) typical, for crest factors
up to 3.
Voltage (1 ms PEAK HOLD)
Range Resolution Accuracy Overload Protection
4 V 0.001 V
1.0 % +43 1000 V R.M.S
40 V 0.01 V
400 V 0.1 V
1000 V 1 V
● Specifi ed accuracy for changes >1 ms in duration
DIODE CHECK/ AUDIBLE CONTINUITY TEST
Range Resolution Accuracy Test Current Open Voltage
Diode 0.001 V 0.5 %+2 Approx. 0.8 mA <+3.1 V
● Overload protection: 1000 V R.M.S. for circuits <0.3 A of short circuit
current.
● Built-in buzzer sounds when reading is below 50 mV approx. and single
tone for normal forward-biased diode or semiconductor junction as 0.3 V
Reading ≤0.8 V.
AC Voltage
Range Resolution Accuracy Overload
Protection
45~400 Hz 400 Hz~2 kHz
4 V 0.001 V
1.0 %+5 2.0 %+5 1000 V R.M.S
40 V 0.01 V
400 V 0.1 V
1000 V 1 V
Notes:
1. Input Impedance: 10 MW (nominal) in parallel with <100 pF
AC+DC Voltage
Range Resolution Accuracy Overload
Protection
45~400 Hz 400 Hz~2 kHz
4 V 0.001 V
1.5 %+9 2.5 %+9 1000 V R.M.S
40 V 0.01 V
400 V 0.1 V
1000 V 1 V
● Input Impedance: 10 MW (nominal) in parallel with <100 pF
DC Voltage
Range Resolution Accuracy Overload Protection
4 V 0.001 V
0.2 %+5
1000 A R.M.S.
40 V 0.01 V
400 V 0.1 V
1000 V 1 V 0.5 %+3
● Input Impedance: 10 MW (nominal).