4 Characteristics and Specifications
Electrical Specifications
70 U1190A Series User’s Guide
Temperature specifications
Table 4 - 5 Temperature specifications
Thermal type Range Resolution
Accuracy
U1194A
K
–40 °C to 400 °C 0.1 °C 1.0% + 2.0 °C
400 °C to 1200 °C 1.0 °C 1.0% + 2.0 °C
K
–40 °F to 752 °F 0.1 °F 1.0% + 3.6 °F
752 °F to 2192 °F 1.0 °F 1.0% + 3.6 °F
Notes for temperature specifications:
1 This function is only applicable for the U1194A model.
2 The specification above is specified after the multimeter has been left stationary in the same operating environment for
1 hour at least. If the unit is exposed during storage in a high humidity (condensing) environment, ensure that the
multimeter has been in the same operating environment for 2 hours at least.
3 The accuracy does not include the tolerance of the thermocouple probe.
4 Do not allow the temperature sensor to contact a surface that is energized above 30 Vrms or 60 V DC. Such voltages pose
a shock hazard.
5 The temperature calculation is specified according to the safety standards of EN/IEC-60548-1 and NIST175.
6 The accuracy specification assumes the surrounding temperature is stable with ±1 °C. For the surrounding temperature
changes of ±3 °C, the rated accuracy applies after two hours.