2 Making Measurements
Measuring Temperature
46 U1190A Series User’s Guide
Changing the default temperature unit
Follow the steps below to change the temperature unit
between Celsius (°C) or Fahrenheit (°F).
1 Press and hold and while powering on the clamp
meter to enter the temperature unit selection menu.
2 Press to make changes to the temperature unit.
Press or to change the temperature unit
(°C or °F).
3 Press to save the changes.
4 Cycle the clamp meter’s power to exit the temperature
unit selection menu.
CAUTION
Always set the temperature unit display per the official requirements
and in compliance with the national laws of your region.
NOTE
The bead-type thermocouple probe is suitable for measuring temperatures
from –40 °C to 204 °C (399 °F) in Teflon-compatible environments. Do not
immerse this thermocouple probe in any liquid. For best results, use a
thermocouple probe designed for each specific application — an
immersion probe for liquid or gel, and an air probe for air measurement.
Observe the following measurement techniques:
• Clean the surface to be measured, and ensure that the probe is
securely touching the surface. Remember to disable the applied power.
• When measuring above ambient temperatures, move the thermocouple
along the surface until you get the highest temperature reading.
• When measuring below ambient temperatures, move the thermocouple
along the surface until you get the lowest temperature reading.
• Place the clamp meter in the operating environment for at least 1 hour
as the clamp meter is using a non-compensation transfer adapter with
miniature thermal probe.