User manual
3 E5810B Web Access
Other Web Access Functions
80 E5810B User’s Guide
2 View the system log from Telnet.
To view the system log from Telnet, click Start and select
Run..., or right- click the bottom- left corner of the Desktop
and select Run (Windows 8). Type telnet <IP_Address>
on the Run dialog box line, where <IP_Address> is the
IP address displayed on the E5810B front panel display.
Refer to Appendix A, “Using the Telnet Utility” for more
information on the Telnet utility. When the Telnet client is
launched, you can view the system log by typing
syslog-display. After a successful firmware update,
the E5810B systems log should show normal startup
messages as shown below.
(00:00:00.00) Selftest Passed
(00:00:00.00) Searching for DHCP Server
(00:00:00.00) GPIB initialized with name : [gpib0]
(00:00:00.00) RS232 initialized with name : [COM1]
(00:00:00.00) USB initialized with name : [usb0]
3 If the update is unsuccessful.
If the last message in the system log indicates the E5810B
was trying to update firmware, the firmware process was
not successful and the E5810B will display a message
dialog. In this case, try the following:
i Verify the FTP server address.
Verify that the firmware FTP server domain name is
correct and fully qualified. A domain name is a
symbolic name for an IP address. For example,
ftp.agilent.com is the domain name for IP Address
192.6.143.21. Try to manually use the FTP to access the
FTP server that contains the E5810B firmware. Verify
that the firmware file is in the proper directory and
that the firmware file has the correct name.