Datasheet
Keysight E3640A – E3649A Programmable DC Power Supply Specications
Model number E3640A E3641A E3642A E3643A E3644A E3645A
Maximum power 30 W 50 W 80 W
# of outputs 1 1 1 1 1 1
DC output rating
(@ 0 ºC to 40 ºC)
0 to 8 V/3 A or
0 to 20 V/1.5 A
0 to 35 V/0.8 A or
0 to 60 V/0.5 A
0 to 8 V/5 A or
to 20 V/2.5 A
0 to 35 V/1.4 A or
0 to 60 V/0.8 A
0 to 8 V/8 A or
0 to 20 V/4 A
0 to 35 V/2.2 A or
0 to 60 V/1.3 A
Net weight 5.3 kg 5.2 kg 6.3 kg 6.2 kg 6.6 kg 6.7 kg
Dimensions (without bumper) 212.6 mm W x 88.5 mm H x 348.3 mm D
Model number E3646A E36 47A E3648A E3649A
Maximum power 60 W 100 W
# of outputs 2 2 2 2
DC output rating
(@ 0 ºC to 40 ºC)
Two 0 to 8 V/3 A or
0 to 20 V/1.5 A
Two 0 to 35 V/0.8 A or
0 to 60 V/0.5 A
Two 0 to 8 V/5 A or
0 to 20 V/2.5 A
Two 0 to 35 V/1.4 A or
0 to 60 V/0.8 A
Net weight 8.2 kg 8.0 kg 9.2 kg 9.1 kg
Dimensions (without bumper) 212.8 mm W x 133.0 mm H x 348.3 mm D
Load and line regulation ± (% of output + offset)
Voltage < 0.01% + 3 mV
Current < 0.01% + 250 μA
Ripple and noise (20 Hz to 20 MHz)
Normal mode voltage < 5 mVpp/0.5 mVrms for 8 V/20 V models
< 8 mVpp/1 mVrms for 35 V/60 V models
Normal mode current < 4 mArms
Common mode current < 1.5 μArms
Accuracy
1
12 months (@ 25 ºC ± 5 ºC), ± (% output + offset)
Programming
Voltage < 0.05% + 10 mV (< 0.1% + 25 mV for output 2 of E3646/47/48/49 A)
Current < 0.2% + 10 mA
Readback (over GPIB and with respect to actual output)
Voltage < 0.05% + 5 mV (< 0.1% + 25 mV for output 2 of E3646/47/48/49 A)
Current < 0.15% + 5 mA (< 0.15% + 10 mA for output 2 of E3646/47/48/49 A)
Meter
2
(over front panel with respect to actual output)
Voltage < 0.05% + 2 counts (<0.1% + 4 counts for output 2)
Current < 0.15% + 5 mA (<0.15% + 10 mA for output 2)
Resolution
Program < 5 mV/1 mA
Readback < 2 mV/1 mA
Meter 10 mV/1 mA
Transient response Less than 50 μsec for output to recover to within 15 mV following a change in output current from full
load to half load or vice versa.
Settling time
3
< 90 msec
OVP
Accuracy,
± (% output + offset)
Activation time
4
< 0.5% + 0.5 V
< 1.5 msec, OVP ¥3 V/< 10 msec, OVP < 3 V
1. Accuracy specications are valid after a 1-hour warm-up with no load and calibration at 25 ºC.
2. Meter accuracy specication is at minimum 10 mV decimal limited by front panel resolution.
3. Maximum time required for the output voltage to change from 1% to 99% or vice versa following the receipt of VOLTage or APPLy command via direct GPIB
or RS-232 interface.
4. Average time for output to start and drop after an OVP condition occurred.
04 | Keysight | E3640A – E3649A Programmable DC Power Supplies – Data Sheet