Chapter 3 Front-Panel Operation
Constant Current Operation
27
3
Limit
Limit
OFF 8V 25V 20V 50V OVP
OCP CC
OVP OCP
CC CV
Note During actual CC operation, if a load change causes the voltage limit to be exceeded,
the power supply will automatically crossover to constant voltage mode at the preset
voltage limit and the output current will drop proportionately.
VOLTage {<voltage>|MIN|MAX} Set the voltage
CURRent {<current>|MIN|MAX} Set the current
OUTPut ON Enable the output
1
Voltage
Current
Display
Limit
Display
Limit
Output
On/Off
*For HP E3633A Model **For HP E3634A Model