Chapter 4 Remote Interface Reference
The SCPI Status Registers
97
4
To Determine When a Command Sequence is Completed
Send a device clear message to clear the power supply’s output buffer (e.g.,
CLEAR 705).
Clear the event registers with the *CLS (clear status) command.
Enable the “operation complete” bit (bit 0) in the Standard Event register by
executing the
*ESE 1 command.
Send the *OPC? (operation complete query) command and enter the result to
ensure synchronization.
Execute your command string to program the desired configuration, and then
execute the
*OPC (operation complete) command as the last command. When
the command sequence is completed, the “operation complete” bit (bit 0) is
set in the Standard Event register.
Use a serial poll to check to see when bit 5 (standard event) is set in the Status
Byte summary register. You could also configure the power supply for an
SRQ interrupt by sending
*SRE 32 (Status Byte enable register, bit 5).
Using *OPC to Signal When Data is in the Output Buffer
*OPC
*OPC
*OPC