Chapter 4 Remote Interface Reference
The SCPI Status Registers
95
4
The Status Byte Summary register is cleared when:
*CLS
*ESR?
The Status Byte Enable register (Request Service) is cleared when:
*SRE 0
*PSC 1
*PSC 0
Using Service Request (SRQ) and Serial POLL
*SRE
Caution The IEEE-488 standard does not ensure synchronization between your bus controller
program and the instrument. Use the *OPC? command to guarantee that commands
previously sent to the instrument have completed. Executing a serial poll before a
*RST, *CLS, or other commands have completed can cause previous conditions
to be reported.