Manual

3
SR-750_AFM_E
COOP2of5
Quiet zone scale factor (4-11)
.......Set the quiet zone scale factor.
CODE39(Trioptic CODE39)
Trioptic CODE39......................Check this to allow reading of Trioptic CODE39.
3 Details of Japanese Pharmaceutical
Code Quality Verification Function
Symbology for code quality verification
•GS1-128
GS1 DataBar Limited
GS1 DataBar Stacked
CC-A (GS1 DataBar Limited)
CC-A (GS1 DataBar Stacked)
Settings of the Japanese Pharmaceutical Code quality
verification function
To use the Japanese Pharmaceutical Code quality verification function, you need
to enable the verification settings.
(1) Setting using the AutoID Network Navigator
Press the "Code quality verification" button on the "Operation" tab in the
Settings view, and then check "Japanese Pharmaceuticals". In addition, select
the code to be verified.
Name of verification function Explanation
Japanese Pharmaceutical Code quality
verification
Function for evaluating the print quality of codes on
pharmaceuticals, recommended by the Safety
Division of Pharmaceutical and Food Safety Bureau
and the Economic Affairs Division of Health Policy
Bureau under the Ministry of Health, Labour and
Welfare.
It makes reference to ISO/IEC15415 and ISO/
IEC15416.
(2) Setting using the setup commands
Send commands that enable the Japanese Pharmaceutical Code quality
verification function.
For configuration commands, refer to "5 Modified and Added Configuration
Commands" (Page 6).
Point
When the Code quality verification function is enabled, the decode
time becomes longer than when it is disabled.
Perform calibration after enabling the Code quality verification
function or set the decode timeout period required for reading.
Use the tact measurement test mode to confirm that there is no
problem with the reading time for operation.
Other settings, functions, and precautions regarding the
code quality verification
The following settings, functions, and precautions are similar to those of ISO/
IEC15415.
For details, refer to "3-11 Code quality verification function in SR-750 Series
User's Manual".
Append Code quality verification result data
For the data format with data appended, refer to "Other Modifications" (Page
7).
Code quality verification function Evaluation standard
Functions that can use evaluation results of the Code quality verification function
Notes when using the Code quality verification function
Description of verification items
Japanese Pharmaceutical Code quality verification
(GS1 DataBar Limited, GS1 DataBar Stacked, GS1-128)
Decode success/failure DEC (Decode)
Description Evaluates whether decoding is possible or not.
-
Calculation
formula
-
Criterion
A : Success
F : Failure
Number of edges EDGE (EdgeDetermination)
Description
Determines whether the read number of edges
is equal to the assumed number of edges.
-
Calculation
formula
-
Criterion
A:Match
F:Mismatch
Symbol contrast SC (Symbol Contrast)
Description
Evaluates the difference between the
maximum brightness value (Rmax) and
minimum brightness value (Rmin) in the code
area.
-
Calculation
formula
SC = Rmax – Rmin
Criterion
A:SC 70 %
B:SC 55 %
C:SC 40 %
D:SC 20 %
F:SC < 20 %
Minimum Reflectance
MINR
(Minimum Reflectance)
Description
Minimum reflectance among scanned
waveforms
-
Calculation
formula
-
Criterion
A:Rmin 0.5 Rmax
F : Rmin > 0.5 Rmax