User Manual

4-5 System Settings (SYS)
4
Settings for Advanced Functions
4-30
- Digital Laser Sensor LV-N10 Series User's Manual -
Received light intensity can be increased 4 times without sacrificing response time.
The maximum received light intensity value which can be displayed differs according
to the power mode setting and display screen.
The maximum received light intensity values which can be displayed are shown below.
"Power Modes" (page 4-4)
"Extension display" (page 4-23)
This function is not supported in the hSP (HIGH SPEED) mode and
the light intensity expansion setting screen is not displayed.
When the preset function is used in DATUM1 or DATUM2 mode,
correctible values are limited to 200.
This cannot be selected when LV-S31 is connected.
Display Gain
Refer to page 4-2 "System Settings (SYS)" for setting methods.
Display Function
Std (Standard) Normal light intensity display (Default)
FuLL (Full) Full light intensity display (Hysteresis increases 4 times)
Display
screen
Display
gain
HIGH
SPEED
FINE TURBO SUPER ULTRA MEGA
Normal dis-
play
Std 4095 8190 9999 9999 9999 9999
FuLL 4095 9999 9999 9999 9999 9999
StdE
(Extension
display)
Std 4095 8190 16380 32760 65520 65472
FuLL - 32760 65520 65504 65472 65280
Point