Owner manual
5
5-23
LS-Navigator-M-NO5-E
5-10 Setting the Measurement Conditions for the Program
1 
Click on the [Calibration] tab in the 
program setting area.
The [Calibration] window appears.
2 
Set [Logic calibration] for area 1 and area 
2.
Note
These settings are effective when [Calibration 
Settings ] for the controller is set to logic calibration 
mode.
For more details, see the “Calibration Settings” 
section in the “User’s Manual” for each controller.
Calibration
Select the area (area 1 or 
area 2).
Enter the values for two points (target 1 and target 
2) on the workpiece.
The values are displayed as follows:
The value for calibration span is automatically 
calculated after the above values are entered.
The calibration span is display
ed as a ref
erence 
value.
Note
When setting the calibration mode, enter values 
that meet the following relationship:
T1-A < T2-A and T1-B < T2-B
Set the values so that the calibration span is 
greater than 0.5 and smaller than 2.0.
Before calibration After calibration
FirstT1-A: Displayed 
value for target 1 
before calibration
T1-B: Displayed 
value for target 1 
after calibration
Second T2-A: Displayed 
value for target 2 
before calibration
T2-B: Displayed 
value for target 2 
after calibration
Copies the calibration settings 
from area 1 or area 2 and pastes 
them into area 1 or area 2. 
1










